DAILY QUALITY CONTROL
11.
DMS
As soon as the screen displays results for the
LOW control, change the Control Run file to
the NORMAL for the lot number you are
using.
Press
and select.
Repeat the selecting and cycling procedure
for HIGH.
12. To review control files and graphs, see CBC
CONTROL FILE REVIEW in the
CONTROLS chapter.
13. If the control is outside limits, the system
•
displays an error message
•
stops if AutoStop is on (status line
shows QC
↑
)
•
turns QC
↑
red
•
flags the result with an H or L.
DMS
to clear the error message before you
continue.
Check Table 3 for possible causes of errors.
See the CONTROLS chapter for these related topics:
•
AutoStop Option
•
Control Management by Shift
•
Error Messages for Controls
•
Control Run Matrix
PN 4237188A (December 1993)