STARTUP
•
If so, did laboratory establish and
enter new ramp and precision
values?
•
If not, you need to do so.
c.
Check that clean diluent covers the
apertures, and that there is no bleach or
cleaning agent in the baths.
d.
Repeat the ramp or precision test.
Press keys next to
ANALYZER
ANALYZER FUNCTIONS
START UP TESTS
RAMP TEST
or
PRECISION TEST
e.
If you see a flag this time,
1
2
3
4
5
6
7
8
9
0
#
*
IN THE USA
COULTER
CUSTOMER OPERATIONS
1-800-526-7694
7188032A
5.
Print a copy on the Graphic
DMS
Printer. Record results in
your logbook.
18
PN 4237188A (December 1993)