CYCLING SAMPLES
6.
Check for
ANALYZER
•
STATUS: READY
•
ALERT: (blank)
DMS
•
Sample Run screen
7.
DILUTER
START/
CONT
8.
Verify all sample ID numbers before you
report results.
7188096A
Secondary Mode: CBC, CBC + DIFF
Use the Secondary mode for manual aspiration. Bar-code labels are not read in this mode; you must
enter an ID number via the Diluter keypad before cycling the sample. You can use only number
characters to identify Secondary mode specimens.
IMPORTANT
Blood detectors are inactive in the Secondary mode. Inspect specimens for clots and use good
laboratory practice for verifying results.
PN 4237188A (December 1993)
61