
STARTUP
12. To review control files and graphs, see
CBC/DIFF CONTROL FILE REVIEW in the
CONTROLS chapter.
13. If a control is outside limits, the system
•
displays an error message
•
stops if AutoStop is on (status line
shows QC
↑
)
•
turns QC
↑
red
•
flags the result with an H or L.
DMS
to clear the error message before you
continue.
Check Table 3 for possible causes of errors.
5C Cell Control for CBC/Diff in the Secondary Mode
5C cell control is assayed only for the Primary mode.
If you run it in the Secondary mode, your
laboratory must
•
determine its own means and expected ranges for each parameter, and
•
set up separate files with separate lot numbers for Secondary mode use.
See Control File Set Up in the Controls chapter for entering mean and expected ranges.
1.
At the Main Menu press
Other
.
DMS
Prime
to turn Prime on.
.
Sample Analysis
Controls
Start Up
Special Functions
DMS
2A
Other
LTER
F3 AR3 Barcode : On
STKS
F5 AutoPrint
: NONE
F6 AutoTransmit: Off
F7 Prime mode
: On
Operator ID
: OPR
ooooo
ooooo
oooooooooooo
oooooooooooo
ooooooooo
CBC/Diff/Retic
(932231) COPYRIGHT 1989-1993 COULTER CORPORATION, ALL RIGHTS RESERVED
F1-Help
F3-Error Log
F5-Other
F9-Main menu
F10-Previous menu
08/20/93 12:53
OPR
DMS
↑
PR
↓
TK
↓
HC
↓
DB
↑
XB
↑
WL
↓
HWL
↓
QC
↑
54
38
PN 4237188A (December 1993)
Summary of Contents for STKS
Page 16: ...CONTROLS AND INDICATORS 6 PN 4237188A December 1993...
Page 84: ...SAMPLE ANALYSIS 74 PN 4237188A December 1993...
Page 98: ...DATA ANALYSIS 88 PN 4237188A December 1993...
Page 124: ...DMS BASICS 114 PN 4237188A December 1993...
Page 130: ...RUN SAMPLES DISPLAY 120 PN 4237188A December 1993...
Page 168: ...DATA BASE 158 PN 4237188A December 1993...
Page 204: ...APPENDIX A 194 PN 4237188A December 1993...
Page 216: ...APPENDIX C 206 PN 4237188A December 1993...
Page 228: ...APPENDIX D 218 PN 4237188B March 1995...
Page 230: ...REFERENCES 220 PN 4237188B March 1995...
Page 236: ......