CYCLING SAMPLES
Secondary mode
Approximately 15 minutes after the last CBC or CBC/Diff sample is run, and no sample has been run
in the Primary mode.
The drain and rinse routine consists of these functions:
•
The Hgb cuvette drains into the waste chamber.
•
The RBC bath drains into the waste chamber.
•
The WBC bath drains into the Hgb cuvette.
•
Diluent refills the baths.
PN 4237188A (December 1993)
73