CYCLING SAMPLES
Secondary Mode: Reticulocytes
IMPORTANT
Perform the reticulocyte dilution method according to the instructions in the package insert and
below. Modifications to the procedure, or failure to follow these instructions, may lead to misleading
or erroneous results.
Prepare the sample using the COULTER ReticPrep reagent kit.
1.
ANALYZER
Set the MODE OF OPERATION to RETICS
so that
CBC
↓
DIFF
↓
RETICS
↑
.
MAIN
SYSTEM CONFIGURATION
MODE OF OPERATION
Press the key opposite the test
configuration to set to RETICS.
MAIN MENU
or
RETURN
*SYSTEM RUN*
STATUS:
ALERT:
READY (RETICS MUST BE DILUTED)
NUMBER OF ASPIRATIONS/TUBE 01
(MESSAGE AREA)
MAIN
03/14/93
10:45:37
BD
CBC
DIFF
RETICS
7188083A
PN 4237188A (December 1993)
67