DAILY QUALITY CONTROL
18. When results appear, check for H (High) or
L (Low) beside the results and for error
messages at the bottom of the screen.
If results are out of limits, see Table 2.
DMS
Use
and
to toggle between DIFF
and RETIC.
DIFF LATEX CONTROL RUN
FILE:
3
LATRON
RUN:
4
EXP DATE: 03/21/94
LOT: 107359
10/10/93
07:01:34
Mean Channel
RESULTS
ASSAY
DIFF
RANGE
+/-
Volume (V)
27.1
27.0
0.1
2.0
Conductivity (C)
26.1
27.0
-0.9
2.0
Scatter (S)
90.9
90.0
0.9
5.0
% CV
RESULTS
EXPECTED CV
<
Volume (V)
3.6
7.0
Diff+Retic
Conductivity (C)
5.1
10.0
REMEMBER TO RUN
|
Scatter (S)
3.2
9.0
IN FUNCTION 57
|
F1-Help F2-File F3-Primer F4-Prt F5-Histo F6-Graph F7-Next F8-Del F9-Menu
↔
Next
10/10/93 07 05
OPR
DMS
↑
PR
↑
TK
↓
HC
↑
DB
↑
XB
↑
WL
↓
HWL
↓
QC
↑
38
19. Press
to print the results for your
logbook.
20. If you want to see graphs of the last 10
cycles, press
Graph
.
To switch between Mean Channel and %CV
graphs, press
Graph
again.
21.
DILUTER
STOP
clears the function showing.
To review the control files and graphs, see
LATEX CONTROL FILE REVIEW in the
CONTROLS chapter.
PN 4237188A (December 1993)
25
Summary of Contents for STKS
Page 16: ...CONTROLS AND INDICATORS 6 PN 4237188A December 1993...
Page 84: ...SAMPLE ANALYSIS 74 PN 4237188A December 1993...
Page 98: ...DATA ANALYSIS 88 PN 4237188A December 1993...
Page 124: ...DMS BASICS 114 PN 4237188A December 1993...
Page 130: ...RUN SAMPLES DISPLAY 120 PN 4237188A December 1993...
Page 168: ...DATA BASE 158 PN 4237188A December 1993...
Page 204: ...APPENDIX A 194 PN 4237188A December 1993...
Page 216: ...APPENDIX C 206 PN 4237188A December 1993...
Page 228: ...APPENDIX D 218 PN 4237188B March 1995...
Page 230: ...REFERENCES 220 PN 4237188B March 1995...
Page 236: ......