WORKFLOW
DMS
COLLATION
CBC/DIFF + RETICS
?
YES
7188084A
PRE-ASSIGNMENT
?
YES
NO
EDIT CBC OR CBC/DIFF
REPORT W/RETIC RESULT
(MANUAL ENTRY FROM
DATABASE QUERY)
NO
5
CBC/DIFF + RETIC
PROFILE IN
WORKLIST
3
ASSIGN RETICS TO
CBC/DIFF
IN DATABASE
4A
ASSIGN CBC/DIFF
TO RETICS
IN DATABASE
4B
PRE-ASSIGNMENT
?
NO
RUN SAMPLES
1
PRE-ASSIGN AS
SEPARATE PROFILE
IN WORKLIST
(FOR EXAMPLE, #1, #4)
2
YES
Table 4 STKS with Retic Analysis Workflow Scenarios
Scenario
What to do
Benefits
1.
No DMS Collation, No Preassignment
•
Retics, CBC/Diff and CBC are
run and reported
independently.
•
Retic reports only RET% (RBC
count can be added to the
report for auto-calculation of
RET#).
Set Analyzer to desired test mode.
Prime Aperture.
Analyze prime sample.
Identify and run your sample(s).
1.
Simplest method,
good for individual
retics or very small
batches.
2.
Useful if host
collates.
continued
PN 4237188A (December 1993)
49
Summary of Contents for STKS
Page 16: ...CONTROLS AND INDICATORS 6 PN 4237188A December 1993...
Page 84: ...SAMPLE ANALYSIS 74 PN 4237188A December 1993...
Page 98: ...DATA ANALYSIS 88 PN 4237188A December 1993...
Page 124: ...DMS BASICS 114 PN 4237188A December 1993...
Page 130: ...RUN SAMPLES DISPLAY 120 PN 4237188A December 1993...
Page 168: ...DATA BASE 158 PN 4237188A December 1993...
Page 204: ...APPENDIX A 194 PN 4237188A December 1993...
Page 216: ...APPENDIX C 206 PN 4237188A December 1993...
Page 228: ...APPENDIX D 218 PN 4237188B March 1995...
Page 230: ...REFERENCES 220 PN 4237188B March 1995...
Page 236: ......