DATA ANALYSIS
Single Parameter Histograms
Single parameter histograms of volume,
conductivity and light scatter are also available
for Diff and Retics. These are used primarily for
troubleshooting by Service.
RETICULOCYTE SCATTERPLOTS
DF 5 Scatterplot
This is the initial display of Retic analysis. It
derives primarily from light scatter and volume.
The Y axis represents volume. Mature RBCs
appear to the left of the Retic discriminator.
Retics appear to the right of it. WBCs appear high
to the far right.