SAMPLE ANALYSIS
10. Label two glass test tubes for each patient
sample: one for Reagent A and one for
Reagent B.
7170049A
IMPORTANT
Dispensing reagent at an angle changes the
dilution of the preparation.
11. Holding the dropper bottle vertically, place 4
drops of Reagent A into the "A" tube.
A
A
7170043A
12. Dispense 50 µL of well-mixed sample into
the "A" tube. Add directly to the stain and
do not let sample run down side of tube.
A
7170044A
13. Gently mix "A" tube.
A
7170045A
70
PN 4237188A (December 1993)