DAILY QUALITY CONTROL
5C Cell Control for CBC/DIFF Test in the Primary Mode
1.
Check that the number of aspirations is set
to 1.
If not,
NUMBER OF ASPIRATIONS/TUBE
DILUTER
1
ENTER
*SYSTEM RUN*
STATUS:
ALERT:
READY
NUMBER OF ASPIRATIONS/TUBE 01
(MESSAGE AREA)
MAIN
03/14/93
10:45:37
BD
CBC
DIFF
RETICS
7188050A
2.
At the Main Menu, press
DMS
Other
.
Prime
to turn Prime on.
.
Sample Analysis
Controls
Start Up
Special Functions
DMS
2A
Other
LTER
F3 AR3 Barcode : On
STKS
F5 AutoPrint
: NONE
F6 AutoTransmit: Off
F7 Prime mode
: On
Operator ID
: OPR
ooooo
ooooo
oooooooooooo
oooooooooooo
ooooooooo
CBC/Diff/Retic
(932231) COPYRIGHT 1989-1993 COULTER CORPORATION, ALL RIGHTS RESERVED
F1-Help
F3-Error Log
F5-Other
F9-Main menu
F10-Previous menu
08/20/93 12:53
OPR
DMS
↑
PR
↓
TK
↓
HC
↓
DB
↑
XB
↑
WL
↓
HWL
↓
QC
↑
54
3.
To prime, press
DILUTER
PRIME
APERT
4.
Cycle a normal whole-blood sample in the
Primary mode:
a.
Put a tube in the cassette.
b.
Place the cassette in the right-side
loading bay.
c.
DILUTER
START/
CONT
7188040A
PN 4237188A (December 1993)
35
Summary of Contents for STKS
Page 16: ...CONTROLS AND INDICATORS 6 PN 4237188A December 1993...
Page 84: ...SAMPLE ANALYSIS 74 PN 4237188A December 1993...
Page 98: ...DATA ANALYSIS 88 PN 4237188A December 1993...
Page 124: ...DMS BASICS 114 PN 4237188A December 1993...
Page 130: ...RUN SAMPLES DISPLAY 120 PN 4237188A December 1993...
Page 168: ...DATA BASE 158 PN 4237188A December 1993...
Page 204: ...APPENDIX A 194 PN 4237188A December 1993...
Page 216: ...APPENDIX C 206 PN 4237188A December 1993...
Page 228: ...APPENDIX D 218 PN 4237188B March 1995...
Page 230: ...REFERENCES 220 PN 4237188B March 1995...
Page 236: ......