DATA ANALYSIS
Reticulocyte Codes
If any of the following flags occur, review results for the affected parameter(s) according to your
laboratory’s protocol.
Table 6 Retic Codes and Messages
On DMS Display
Cause
Message in Condition Box
::::: FC under Scatterplot
Full clog
VERIFY RETIC
.....
Incomplete.
Not enough events.
VERIFY RETIC
RBC XX.XXE
RET# .XXXXe
E = Operator entered RBC
e = Parameter calculated from
edited or manually entered
RBC or Retic
EDITED DATA
RET# .XXXXE
Operator edited RET#
EDITED DATA
RET% XX.XXE
Operator edited RET%
EDITED DATA
RET% XX.XX+
RET# .XXXXR
RET% >30.0
VERIFY RETIC
RET% XX.XXR
RET# .XXXXR
RET% <0.5 or
Abnormal data pattern
VERIFY RETIC
RET% XX.XXL or XX.XXH
RET# .XXXXL or .XXXXH
Results < or > Action Limits
VERIFY RETIC
RET# +++++
Exceeds limit of display for RET#
when data is edited
EDITED DATA
RET% ?????
Invalid data. RET% <0.00 or
>100.0
VERIFY RETIC
Messages
There are four types of messages:
Population Condition
Definitive
Suspect
Analyzer Transmit
82
PN 4237188A (December 1993)
Summary of Contents for STKS
Page 16: ...CONTROLS AND INDICATORS 6 PN 4237188A December 1993...
Page 84: ...SAMPLE ANALYSIS 74 PN 4237188A December 1993...
Page 98: ...DATA ANALYSIS 88 PN 4237188A December 1993...
Page 124: ...DMS BASICS 114 PN 4237188A December 1993...
Page 130: ...RUN SAMPLES DISPLAY 120 PN 4237188A December 1993...
Page 168: ...DATA BASE 158 PN 4237188A December 1993...
Page 204: ...APPENDIX A 194 PN 4237188A December 1993...
Page 216: ...APPENDIX C 206 PN 4237188A December 1993...
Page 228: ...APPENDIX D 218 PN 4237188B March 1995...
Page 230: ...REFERENCES 220 PN 4237188B March 1995...
Page 236: ......