DAILY QUALITY CONTROL
To review the Retic Control files and graphs,
see RETIC CONTROL FILE REVIEW in the
CONTROLS chapter.
If Retic Control values are out of range, see
Table 3, When a Control is Out of Limits.
21. For the next two levels,
•
Select the corresponding Retic Control
Run files.
•
Cycle the prepared controls.
5C Cell Control for CBC/Diff
5C cell controls have bar-code labels.
When the bar-code label is read, or when you enter the lot number in the Secondary mode properly,
the DMS automatically:
•
Identifies the control by level, lot, expiration date and shift.
•
Checks control expiration date against current date.
•
Stores the data in the proper control file.
•
Compares results to expected ranges.
•
Flags out-of-range results by H (high) or L (low).
•
Displays certain error messages, if they apply, at the bottom of the DMS display.
•
Displays the status of the last control run with a QC flag.
•
Stops the system when Auto-Stop conditions apply.
Cycle 5C cell control in the CBC/Diff test mode.
PN 4237188A (December 1993)
33
Summary of Contents for STKS
Page 16: ...CONTROLS AND INDICATORS 6 PN 4237188A December 1993...
Page 84: ...SAMPLE ANALYSIS 74 PN 4237188A December 1993...
Page 98: ...DATA ANALYSIS 88 PN 4237188A December 1993...
Page 124: ...DMS BASICS 114 PN 4237188A December 1993...
Page 130: ...RUN SAMPLES DISPLAY 120 PN 4237188A December 1993...
Page 168: ...DATA BASE 158 PN 4237188A December 1993...
Page 204: ...APPENDIX A 194 PN 4237188A December 1993...
Page 216: ...APPENDIX C 206 PN 4237188A December 1993...
Page 228: ...APPENDIX D 218 PN 4237188B March 1995...
Page 230: ...REFERENCES 220 PN 4237188B March 1995...
Page 236: ......