2. Installation Manual
- 82 -
(16) Tool Modes (/np mode)
1) Purpose
The tool modes (startup options) are intended in order to check operation, and are used
to launch the CXDI application on the control PC by itself, and to display items that are
not normally displayed.
2) Notes
The following operation must be performed before using the “/np” mode. Especially, be
sure to back up the exposure mode names and the customized settings before the
operation.
2-1) When using “/np” with the same settings as that of the connected sensor unit
“BodyPart**.ini” file can be used as it is.
Example:
Connection
/np
setting
Sensor1 Table
Table
Sensor2 Stand
Stand
2-2) In case of using “/np” with different settings from that of the connected sensor unit
Move the “BodyPart**.ini” file in the “BodyParts” folder to the desktop, etc.
However, do not move the Reference folder. If the “BodyPart**.ini” file is left in the
“BodyParts” folder, system will not be able to be started, as the sensor type of the
“BodyPart” and the settings do not match.
Example:
Connection
/np
setting
Sensor1 Table
Table
Sensor2 Stand
Stand
3) Preparation
3-1) Connect the keyboard and the mouse to the control PC.
3-2) Delete the “ccrstart.bat” file from startup.
3-3) Disconnect the imaging unit from the control PC.
4) Startup
method
4-1) Start up Windows XP.
4-2) Start the Command Prompt screen.
Start
⇒
Program
⇒
Accessories
⇒
Command Prompt
4-3) Command Prompt screen appears, type the commands following instruction below to
start the CXDI application. (Press the [Enter] key after typing the command.)
No. Command
Prompt
Command
Note.
1 C:\>
D:
2 D:\>
cd
ccr
“Space” delimiter is required between
“cd” and “ccr”.
3 D:\ccr>
ccrxxxxx
/np
“Space” delimiter is required between
“xxxxx” and “/”.
“xxxxx” is different in version.
4-4) If the message “Sensor not connected” appears at starting of the CXDI application,
click [OK] button, Change to the “Debugging mode” with the keys ([Alt] + [Tab])
using.
4-5) When Welcome to CCR appears, select “1. Set-Up...”.
Summary of Contents for CXDI-50G
Page 20: ...2 Installation Manual 3 CXDI 50G Power Box assemble package 4 Grid optional 6 ...
Page 32: ...2 Installation Manual When normal imaging When timeout due to RX_REQ not negating 18 ...
Page 58: ...2 Installation Manual 44 Input parameters Fig 3 ...
Page 104: ...2 Installation Manual 90 2 Power Box Unit mm Mass 4 2Kg ...