2. Installation Manual
The connection with CXDI-50G and X-ray generator equipment
Connecting condition
(1)
The X-ray emission signal line impedance is below 100
Ω
.
(2)
The max voltage at the point of X-ray release signal line is AC250V max, DC30V max
and its current is from 10mA to 2A. However only 2
nd
Power source side is available for
connecting.
- 20 -