ror Code List for Versi
Error Code List for Version 6.00
Error Code
Seriousness
Reply
Keys
Message
[Cause]
[Operation after Reply Key is touched]
-5101
Error
OK #1:Error
#2:Sensor Unit:Detect Error
#3:Check the sensor unit cable connection. If you want to
recover the system, shut it down.
[Cause]
The cable is not connected though it is setting which uses the
sensor.
[Operation after Reply Key is touched]
The system only requires your confirmation. Touch OK. Only
the problematic sensor is disabled.
-5102
Error
OK #1:Error
#2:Grid Error
#3:Remove Grid for this exposure.
[Cause]
This error occurs when grids are detected in the
Buck/Cassette setting, where imaging is expected to be
performed without grids.
[Operation after Reply Key is touched]
Touch OK after the check.