![Canon CXDI-50G Service Manual Download Page 167](http://html.mh-extra.com/html/canon/cxdi-50g/cxdi-50g_service-manual_61474167.webp)
ror Code List for Versi
Error Code List for Version 6.00
Error Code
Seriousness
Reply
Keys
Message
[Cause]
[Operation after Reply Key is touched]
-509
Error
OK #1:Error
#2:External Storage Error:
#3:Selected directory or disk for storage can only be read.
[Cause]
The external storage medium for the stored data is read-only.
[Operation after Reply Key is touched]
The system returns to the state before the occurrence of the
error.
-512
Error
OK #1:Error
#2:External Storage Error:
#3:Cannot recognize the removable disk. Set the disk properly
and retry.
[Cause]
No medium is found in the removable disk.
[Operation after Reply Key is touched]
The system returns to the state before the occurrence of the
error.
-513
Error
OK #1:Error
#2:Media Control Error
#3:Operation has been canceled because the external storage
or directory is busy.
[Cause]
The external medium is busy deleting or writing the data. This
error occurs such as when you try to eject the disk while the
external medium is busy.
[Operation after Reply Key is touched]
After you confirm it, touch OK.
-514
Error
OK #1:Error
#2:External Storage Error:
#3:The exposed images cannot be stored because the size of
folder partition is smaller than the exposed images. Use a new
disk, erase the folder contents or enlarge the folder size and
retry.
[Cause]
An image cannot be stored into a folder in the external storage
medium when the size of the folder is smaller than the image.
Enlarge the folder size in the system setting.
[Operation after Reply Key is touched]
When RETRY is chosen, the image is stored again, while
selecting CANCEL cancels storing the image to the external
storage medium. When CANCEL is chosen, a confirmation
message appears before deleting the image.
- 12 -
Summary of Contents for CXDI-50G
Page 20: ...2 Installation Manual 3 CXDI 50G Power Box assemble package 4 Grid optional 6 ...
Page 32: ...2 Installation Manual When normal imaging When timeout due to RX_REQ not negating 18 ...
Page 58: ...2 Installation Manual 44 Input parameters Fig 3 ...
Page 104: ...2 Installation Manual 90 2 Power Box Unit mm Mass 4 2Kg ...