ror Code List for Versi
Error Code List for Version 6.00
Error Code
Seriousness
Reply
Keys
Message
[Cause]
[Operation after Reply Key is touched]
-311
Error
OK #1:Error
#2:REX Value Warning
#3:REX is 9999. It is under the maximum limitation value
999.
Please reconfirm the generator setting or exposure
condition.
If the condition is correct, adjust the image processing.
[Cause]
This error occurs when the REX value acquired by image
processing after the imaging is more than the limitation value.
[Operation after Reply Key is touched]
Nothing
[Note]
9999: REX value obtained by image processing
999: upper limit value
-402
Error
OK #1:Error
#2:Incomplete End Error
#3:The last image in the last study was not saved completely.
[Cause]
Failed to write in the last image of the previous imaging.
[Operation after Reply Key is touched]
If there remain the previous images, they will be recovered.
-403
Error
OK #1:Error
#2:Incomplete End Error
#3:Do you want store captured images into internal storage in
order to recover them later?
(Name of the patient)
(Number of images)
[Cause]
The last image(s) was not queued due to incomplete end in
the previous imaging.
[Operation after Reply Key is touched]
When Ok is chosen, the recovery process will start. When
Cancel is chosen, the last images will be deleted. In this case,
normal exposure for imaging will start without recovering the
last images. When CANCEL is chosen, a confirmation
message will appear before deleting the image.
- 11 -
Summary of Contents for CXDI-50G
Page 20: ...2 Installation Manual 3 CXDI 50G Power Box assemble package 4 Grid optional 6 ...
Page 32: ...2 Installation Manual When normal imaging When timeout due to RX_REQ not negating 18 ...
Page 58: ...2 Installation Manual 44 Input parameters Fig 3 ...
Page 104: ...2 Installation Manual 90 2 Power Box Unit mm Mass 4 2Kg ...