![Canon CXDI-50G Service Manual Download Page 165](http://html.mh-extra.com/html/canon/cxdi-50g/cxdi-50g_service-manual_61474165.webp)
ror Code List for Versi
Error Code List for Version 6.00
Error Code
Seriousness
Reply
Keys
Message
[Cause]
[Operation after Reply Key is touched]
-305
Error
OK #1:Error
#2:Calibration Error
#3:Collimator may be used.
[Cause]
This error is not currently used.
[Operation after Reply Key is touched]
Perform imaging again.
-306
Error
OK #1:Error
#2:QA Process Error
#3:Image Analysis Error. Adjust image process parameter
manually.
[Cause]
Feature extraction may have failed.
[Operation after Reply Key is touched]
The user must process the image manually.
-310
Error
OK #1:Error
#2:REX Value Warning
#2:REX Value Warning
#3:REX is 999. It is under the minimum limitation value
9999.
Please reconfirm the generator setting or exposure
condition.
If the condition is correct, adjust the image processing.
[Cause]
This error occurs when the REX value acquired by image
processing after the imaging is less than the limitation value.
[Operation after Reply Key is touched]
Nothing
[Note]
999: REX value obtained by image processing
9999: lower limit value
- 10 -
Summary of Contents for CXDI-50G
Page 20: ...2 Installation Manual 3 CXDI 50G Power Box assemble package 4 Grid optional 6 ...
Page 32: ...2 Installation Manual When normal imaging When timeout due to RX_REQ not negating 18 ...
Page 58: ...2 Installation Manual 44 Input parameters Fig 3 ...
Page 104: ...2 Installation Manual 90 2 Power Box Unit mm Mass 4 2Kg ...