![Canon CXDI-50G Service Manual Download Page 159](http://html.mh-extra.com/html/canon/cxdi-50g/cxdi-50g_service-manual_61474159.webp)
ror Code List for Versi
Error Code List for Version 6.00
Error Code
Seriousness
Reply
Keys
Message
[Cause]
[Operation after Reply Key is touched]
-6
Alert
OK #1:Alert
#2:System Info Error
#3:A/D board info is updated.
[Cause]
The A/D board information is different from the former
information. This problem occurs when the version of the
A/D board has been updated or when the connected units have
been changed. If this error occurs when no such changes have
been made, it is attributed to another problem. In this case, a
service person needs to be called in.
[Operation after Reply Key is touched]
The A/D board information is renewed.
-8
Alert
OK #1:Alert
#2:System SelfCheck Error
#3:Call a service person and let him/her know the error code.
[Cause]
An error has been found by a self-check.
[Operation after Reply Key is touched]
Although imaging is possible, image may be degraded.
Therefore, a service person needs to be called in.
-116
Alert
OK #1:Alert
#2:Sensor Unit:SelfCheck Error
#3:Detected an Error.
#4:ER(n) (xxxx)
n: Sensor Index (1 to 4)
xxxx: Error code
[Cause]
An undefined error has been found in the sensor A/D board.
[Operation after Reply Key is touched]
Nothing
-5002
Alert
OK #1:Alert
#2:OPU-CCR RPC Error
#3:OPU-CCR RPC Error.
[Cause]
For some reason, the CCR software has shut down.
[Operation after Reply Key is touched]
- 4 -
Summary of Contents for CXDI-50G
Page 20: ...2 Installation Manual 3 CXDI 50G Power Box assemble package 4 Grid optional 6 ...
Page 32: ...2 Installation Manual When normal imaging When timeout due to RX_REQ not negating 18 ...
Page 58: ...2 Installation Manual 44 Input parameters Fig 3 ...
Page 104: ...2 Installation Manual 90 2 Power Box Unit mm Mass 4 2Kg ...