![Canon CXDI-50G Service Manual Download Page 173](http://html.mh-extra.com/html/canon/cxdi-50g/cxdi-50g_service-manual_61474173.webp)
ror Code List for Versi
Error Code List for Version 6.00
Error Code
Seriousness
Reply
Keys
Message
[Cause]
[Operation after Reply Key is touched]
-504
Warning
RETRY
CANCEL
#1:Warning
#2:DICOM Parameter Error
#3:DICOM transfer Parameter is wrong.
[Cause]
DICOM transfer parameters are not correct.
[Operation after Reply Key is touched]
When RETRY is chosen, the image will be transferred again.
When CANCEL is chosen, transfer will be canceled after
displaying a confirmation message. Enter the correct
parameters and transfer the image from the study list again.
-505
Warning
RETRY
CANCEL
#1:Warning
#2:DICOM Response Error
#3:Receiving device reported an error in DICOM response.
Check receiving device.
[Cause]
DICOM response was received properly. However, the target
reported an error as a response.
[Operation after Reply Key is touched]
When RETRY is chosen, the image will be transferred again.
When CANCEL is chosen, transfer will be canceled after
displaying a confirmation message. Re-transfer is possible by
sending an image from the study list.
-507
Warning
RETRY
CANCEL
#1:Warning
#2:DICOM Printer Status Error
#3:CXDI has aborted DICOM transmission because of a
FAILURE in DICOM printer status. Check receiving device.
[Cause]
DICOM printer status was received properly. However, the
target reported a FAILURE as printer status, and so the image
transfer was cancelled.
[Operation after Reply Key is touched]
When RETRY is chosen, the image will be transferred again.
When CANCEL is chosen, the transfer will be cancelled after
displaying a confirmation message. Re-transfer is possible by
sending the image from the study list.
- 18 -
Summary of Contents for CXDI-50G
Page 20: ...2 Installation Manual 3 CXDI 50G Power Box assemble package 4 Grid optional 6 ...
Page 32: ...2 Installation Manual When normal imaging When timeout due to RX_REQ not negating 18 ...
Page 58: ...2 Installation Manual 44 Input parameters Fig 3 ...
Page 104: ...2 Installation Manual 90 2 Power Box Unit mm Mass 4 2Kg ...