![Canon CXDI-50G Service Manual Download Page 175](http://html.mh-extra.com/html/canon/cxdi-50g/cxdi-50g_service-manual_61474175.webp)
ror Code List for Versi
Error Code List for Version 6.00
Error Code
Seriousness
Reply
Keys
Message
[Cause]
[Operation after Reply Key is touched]
-514
Warning
RETRY
CANCEL
#1:Warning
#2:External Storage Error:
#3:The exposed images cannot be stored because the size of
folder partition is smaller than the exposed images. Use a new
disk, erase the folder contents or enlarge the folder size and
retry.
(The size of folder partition: xxxxMB)
(The size of exposed images: xxxxMB)
[Cause]
Images cannot be stored into the folder when the folder size
of the external storage medium is smaller than a study.
[Operation after Reply Key is touched]
When RETRY is chosen, the image will be stored.
When CANCEL is chosen, storing of image will be cancelled
after displaying a confirmation message. Re-transfer is
possible by transferring the image from the study list.
-515
Warning
RETRY
CANCEL
#1:Warning
#2:Internal Storage Error:
#3:Selected directory for storage can only be read.
[Cause]
The external storage medium is read-only.
Use a writable external storage medium.
[Operation after Reply Key is touched]
When RETRY is chosen, the image will be stored into the
internal disk.
When CANCEL is chosen, storing of image will be cancelled
after displaying a confirmation message. Re-transfer is
possible by sending the image from the study list.
-516
Warning
RETRY
CANCEL
#1:Warning
#2:Internal Storage Error:
#3:Selected directory for storage does not exist, or attribution
is incorrect.
[Cause]
There is no directory in the destination, or recording cannot
be performed due to system function error.
[Operation after Reply Key is touched]
When RETRY is chosen, the image will be stored into the
internal disk.
When CANCEL is chosen, storing of image will be cancelled
after displaying a confirmation message. Retransfer is
possible by sending the image from the study list.
- 20 -
Summary of Contents for CXDI-50G
Page 20: ...2 Installation Manual 3 CXDI 50G Power Box assemble package 4 Grid optional 6 ...
Page 32: ...2 Installation Manual When normal imaging When timeout due to RX_REQ not negating 18 ...
Page 58: ...2 Installation Manual 44 Input parameters Fig 3 ...
Page 104: ...2 Installation Manual 90 2 Power Box Unit mm Mass 4 2Kg ...