![Canon CXDI-50G Service Manual Download Page 60](http://html.mh-extra.com/html/canon/cxdi-50g/cxdi-50g_service-manual_61474060.webp)
2. Installation Manual
- 46 -
D. After setting the output destinations, follow the procedure below to check
whether images can actually be transmitted. Return to the user menu, capture a
sample image (one X-ray image), and transmit the image to the printer and
storage. There are two errors that may result if the image cannot be transmitted:
a. “DICOM Connect Error. Cannot connect to the target. Check network or port
number setting. Retry?”
b. “DICOM Transfer Error. Error occurred during the association. Retry?”
Message (a) indicates that connection at the TCP/IP level is not possible and that the
physical connections or the subnet mask and other settings must be checked again.
Message (b) indicates that communication at the TCP/IP level is problem-free but
that DICOM level communication has failed. In this case, check again that
AE_TITLE of CXDI has been sent properly to the transmission destination and that
the IP address, port number and AE_TITLE of the transmission destination which are
set with CXDI have been set properly.
* “AE_TITLE” of the transmission destination is case sensitive fields.
(Permit upper-case letter or lower case letter, etc)
Summary of Contents for CXDI-50G
Page 20: ...2 Installation Manual 3 CXDI 50G Power Box assemble package 4 Grid optional 6 ...
Page 32: ...2 Installation Manual When normal imaging When timeout due to RX_REQ not negating 18 ...
Page 58: ...2 Installation Manual 44 Input parameters Fig 3 ...
Page 104: ...2 Installation Manual 90 2 Power Box Unit mm Mass 4 2Kg ...