![Canon CXDI-50G Service Manual Download Page 38](http://html.mh-extra.com/html/canon/cxdi-50g/cxdi-50g_service-manual_61474038.webp)
2. Installation Manual
3.2 Network setting for Screwcap.ini
The CXDI software communicates with the imaging part through screwcap.dll by using the
communication protocol for sending and receiving commands and responses.
In conjunction with the communication, Screwcap.ini retains the information of the connected
CXDI-50G and control PC such as network addresses.
It is necessary to edit Screwcap.ini to communicate with the imaging part.
Since the Screwcap.ini factory setting is the following default setting (see the figure below), it is
not necessary to edit the setting unless you changed the network protocol TCP/IP setting for the
LAN card that communicates with the CXDI-50G imaging part in the previous item, “3.1
TCP/IP setting for the control PC network card”.
Screwcap.ini is located in the following directory:
D:\ccr\screwcap.ini
CXDI-50G IP address.
*1
This must be the same as the
IP address stored in the
imaging part.
[HostInfo]
IpAddress=
192.168.100.10
SubnetMask=255.255.255.0
ResponseTimeout=30000
DataTimeout=60000
IntervalTimeout=30000
[SensorInfo]
SensorNum=
1
[Sensor1]
IpAddress=
192.168.100.11
CommandPort=12121
DataPort=12122
[Log]
LogLevel=2
[SensorInfo]
SensorNum=1 Confirm this
LAN card IP address
*
1
CXDI-50G’s IP address: This address must be the same as the IP address stored in the next
item, “3.3 Network setting stored in the CXDI-50G sensor”.
- 24 -
Summary of Contents for CXDI-50G
Page 20: ...2 Installation Manual 3 CXDI 50G Power Box assemble package 4 Grid optional 6 ...
Page 32: ...2 Installation Manual When normal imaging When timeout due to RX_REQ not negating 18 ...
Page 58: ...2 Installation Manual 44 Input parameters Fig 3 ...
Page 104: ...2 Installation Manual 90 2 Power Box Unit mm Mass 4 2Kg ...