![Canon CXDI-50G Service Manual Download Page 76](http://html.mh-extra.com/html/canon/cxdi-50g/cxdi-50g_service-manual_61474076.webp)
2. Installation Manual
- 62 -
The printer status
has returned a
warning.
Return value: 0x28
CXDI description: DICOM printer status warning
(CCRTRANS_WRN_PRN_STATUS)
The transfer process was successful, but a warning was returned as the printer
status.
The log output in this case is shown below.
------------------------------------------------------------------------------------
(Not determined)
------------------------------------------------------------------------------------
Other errors
Return value: Values other than those above
CXDI description : DICOM communications error
(CCRTRANS_ERR_DICOM_TRANSE)
The log output in this case depends on the specific error. A typical example is
shown below.
------------------------------------------------------------------------------------
18-140933[d2]ERR : c0082 SRV Send (DATA SET) failed in
SRV_SendDataSet
18-140933[d2]ERR : 190082 SRV Request failed in
SRV_NCreateRequest
18-140933[d2]ERR : 70012 NULL_key passed to
routineDUL_ReleaseAssociation
------------------------------------------------------------------------------------
Summary of Contents for CXDI-50G
Page 20: ...2 Installation Manual 3 CXDI 50G Power Box assemble package 4 Grid optional 6 ...
Page 32: ...2 Installation Manual When normal imaging When timeout due to RX_REQ not negating 18 ...
Page 58: ...2 Installation Manual 44 Input parameters Fig 3 ...
Page 104: ...2 Installation Manual 90 2 Power Box Unit mm Mass 4 2Kg ...