Error Return Values and Log Output for print_stuff (Reference)
Error example
Return value and log output
Success Return
value:
0x00
There is no log output in this case.
Invalid parameter
Return value: 0x00000001
CXDI description : DICOM connection error
(CCRTRANS_ERR_DICOMPARAM)
The log output in this case is shown below.
-----------------------------------------------------------------------------------
**
-s
Silent mode; do not print results of all print commands
**
-v
Use verbose mode for DUL and SRV facilities
**
x
Canon Hidden Special Mode
node
The host name that is running a print server
port
TCP/IP port number of print server
file
One or more files that contain preformatted images for printing
------------------------------------------------------------------------------------
The server has not
started up.
Return value: 0x00180012
CXDI description : DICOM connection error
(CCRTRANS_ERR_NOT_CONNECT)
------------------------------------------------------------------------------------
The log output in this case is shown below.
18-135933[d2]ERR:
d0012 Attempt to connect to unknown host: test
18-135933[d2]ERR:
130012 Peer aborted Association
(or
never
connected)
18-135933[d2]ERR:
180012 Failed to establish association
------------------------------------------------------------------------------------
After a command
request was sent to
the server, an error
was returned in
response.
Return value: 0x10
CXDI description: DICOM response error (CCRTRANS_ERR_RESP)
The log output in this case is shown below.
------------------------------------------------------------------------------------
(Not determined)
------------------------------------------------------------------------------------
After a command
request was sent to
the server, a
warning was
returned in
response.
Return value: 0x18
CXDI description :
DICOM response warning
(CCRTRANS_WRN_RESP)
The transfer process was successful, but a warning was returned from the server.
The log output in this case is shown below.
------------------------------------------------------------------------------------
(Not determined)
------------------------------------------------------------------------------------
The printer status
has returned an
error.
Return value: 0x20
CXDI description :
DICOM printer status error
(CCRTRANS_ERR_PRN_STATUS)
The log output in this case is shown below.
------------------------------------------------------------------------------------
(Not determined)
------------------------------------------------------------------------------------
Summary of Contents for CXDI-50G
Page 20: ...2 Installation Manual 3 CXDI 50G Power Box assemble package 4 Grid optional 6 ...
Page 32: ...2 Installation Manual When normal imaging When timeout due to RX_REQ not negating 18 ...
Page 58: ...2 Installation Manual 44 Input parameters Fig 3 ...
Page 104: ...2 Installation Manual 90 2 Power Box Unit mm Mass 4 2Kg ...