![Canon CXDI-50G Service Manual Download Page 42](http://html.mh-extra.com/html/canon/cxdi-50g/cxdi-50g_service-manual_61474042.webp)
2. Installation Manual
2-8) The display returns to the “DENSITY ADJUSTMENT CONTROL” screen.
Confirm that POSITION and SIZE fields not set in step 5) to step 7) are disabled
(dimmed). If they are not dimmed, press the [Area setting] key, and press
[DISABLE] in fixed ROI * settings.
Check to make sure the displays are dimmed.
2-9) After Confirming all settings, and press [EXIT].
2-10) The display returns to the system settings screen. Press [OK].
2-11) “Change settings?” appears. Press [OK]. Be careful, because if [CANCEL] is
pressed, all changes made to the settings are deleted.
“Change settings?” appeared.
2-12) Return to the normal imaging screen, and turn off the power to
the CXDI.