ror Code List for Versi
Error Code List for Version 6.00
Error Code
Seriousness
Reply
Keys
Message
[Cause]
[Operation after Reply Key is touched]
OK
Cancel
#1:Error
#2: Sensor Unit: PLD Error
#3: Detected PLD Error. Call a service person.
[Cause]
A PLD error has been found in the sensor A/D board.
[Operation after Reply Key is touched]
When OK key is touched, the sensor will be disconnected.
(Access will not be made, as the system is in an "abnormal"
state.)
If this error persists even after the system is rebooted, a
service person needs to be called in.
-123
Error
0xeff8
0xeff7
"PLD err (n)( eff8)”
"PLD Config err (n)( eff7)”
n: Sensor Index (1 to 4)
OK #1:Error
#2:Sensor Unit:Grid Error
#3:Because of the Grid error, this image may have a quality
problem.
[Cause]
Although a warning for grid had been detected in the sensor
A/D board, imaging was continued. Thus, the image may
have been degraded.
[Operation after Reply Key is touched]
When OK key is touched, the screen returns to the normal
display without doing anything.
-125
Error
0xeffb
"Grid was changed during exposure (n)(effb)"
n: Sensor Index (1 to 4)
- 7 -
Summary of Contents for CXDI-50G
Page 20: ...2 Installation Manual 3 CXDI 50G Power Box assemble package 4 Grid optional 6 ...
Page 32: ...2 Installation Manual When normal imaging When timeout due to RX_REQ not negating 18 ...
Page 58: ...2 Installation Manual 44 Input parameters Fig 3 ...
Page 104: ...2 Installation Manual 90 2 Power Box Unit mm Mass 4 2Kg ...