2. Installation Manual
- 74 -
<
Reference 2>
Example when the graph does not form a straight line
In the data and graph below, the data was obtained by using the Kodak MLP 190 to print
out the SMPTE test image, and the image data was measured.
(1): Shows normal data.
(2): Shows the case when the CXDI printer parameters are not set linearly.
(3): Shows the case when the printer was not calibrated properly.
SMPTE Density
Step
1
2
3
4
5
6
7
8
9
10
11
Percentage
0
10
20
30
40
50
60
70
80
90
100
0.17 0.46 0.75 1.03 1.36 1.66 1.98 2.29
2.59
2.90
3.20
(2)
Inadequate parameter settings
0.17 0.29 0.45 0.61 0.80 1.02 1.26 1.59
2.02
2.53
3.19
(3) Inadequate
0.17 0.33 0.55 0.81 1.08 1.36 1.66 1.97
2.28
2.59
2.94
(1) Normal
0.00
0.50
1.00
1.50
2.00
2.50
3.00
3.50
1
2
3
4
5
6
7
8
9
10
11
Step
Densi
ty
Normal
Inadequate
setting
Inadequate
* For improper parameters, the printer parameter “-S CS000” was not entered.
* For the inadequate calibration, the printer calibration data was set too low.