![Canon CXDI-50G Service Manual Download Page 66](http://html.mh-extra.com/html/canon/cxdi-50g/cxdi-50g_service-manual_61474066.webp)
2. Installation Manual
- 52 -
-B border_density
Border density (Dx100)
(BLACK/WHITE/D x
100)
* This parameter is used in the
DICOM Basic Film Box (2010,
0100).
* This parameter determines the
area density around the image
on the film.
Æ
If this parameter is not designated,
it is not transferred over DICOM. In
this case, the printer default values are
used.
-G empty_image_density
Empty image density
(BLACK/WHITE/D x
100)
* This parameter is used in the
DICOM Basic Film Box (2010,
0110).
This parameter designates the
density of the empty image
area during multi-formatting.
Æ
If this parameter is not designated,
it is not transferred over DICOM. In
this case, the printer default values are
used.
-T trim
Trimming (NO/YES)
* This parameter is used in the
DICOM Basic Film Box (2010,
0140).
* This parameter adds lines
around the image.
* The CXDI is normally adjusted
so that the trimming does not
appear.
Æ
If this parameter is not designated,
it is not transferred over DICOM. In
this case, the printer default values are
used.
* For example, the default value for
this parameter in Agfa printers is
YES. Therefore, the NO option
needs to be specifically designated if
it is desired.
-P polarity
Polarity
(NORMAL/REVERSE)
* This parameter is used in the
DICOM Basic Image Box
(2020, 0020).
* Reverse image density
Æ
If this parameter is not designated,
it is not transferred over DICOM. In
this case, the printer default values are
used.
-r pixel_pitch
Transfer pixel pitch for
designating the request
image size (Pixel Pitch in
um)
* This parameter is used in the
DICOM Basic Image Box
(2020, 0020).
* Reverse image density
* This parameter is used in the
DICOM Basic Image Box
(2020, 0010).
* Position of the image on film
* The cumulative value for the
horizontal size of the image
(raw) at the designated pixel
pitch is used for the request
image size.
* The CXDI automatically uses
131072 internally for
operation.
* The above value can be
overwritten by designating a
specific value.
Æ
If this parameter is not designated,
it is not transferred over DICOM. In
this case, the type of image that is
printed depends on the settings at the
printer side.
-N annoFmt
Annotation position
(1/2/3)
* This parameter is used in the
DICOM Basic Annotation Box
(2030, 0010).
* This parameter designates the
position of the character string
to be annotated.
* If using annotation, always be sure to
transfer the annotation position.
-n annotation
Annotation (character
string)
* This parameter is used in the
DICOM Basic Annotation Box
(2030, 0020).
* This parameter designates the
character string to be
annotated.
* If this parameter is not designated, it
is not transferred over DICOM. In
this case, the printer default values
are used.
* Also, in this case, the type of image
that is printed depends on the
settings at the printer side.
Summary of Contents for CXDI-50G
Page 20: ...2 Installation Manual 3 CXDI 50G Power Box assemble package 4 Grid optional 6 ...
Page 32: ...2 Installation Manual When normal imaging When timeout due to RX_REQ not negating 18 ...
Page 58: ...2 Installation Manual 44 Input parameters Fig 3 ...
Page 104: ...2 Installation Manual 90 2 Power Box Unit mm Mass 4 2Kg ...