ror Code List for Versi
Error Code List for Version 6.00
Error Code
Seriousness
Reply
Keys
Message
[Cause]
[Operation after Reply Key is touched]
-100
Fatal
OK #1:Fatal
Error
#2:Sensor Unit: SelfCheck Error
#3:Detected an Error.
#4:ER(n) (xxxx)
xxxx: Error code
[Cause]
An undefined error has been found in the sensor A/D board.
[Operation after Reply Key is touched]
Nothing
-14
Error in starting the task in DLL (Failed to begin thread.)
-15
Error in DLL resource (Driver does not start.)
-95
Number of sensor units is incorrect (No sensor unit is
connected.)
-102
Fatal
-98
Capture device cannot be found.
-103
Fatal
Shut Down
#1:Fatal Error
#2:READ Timeout Error
#3:Failed to transmit the data.
[Cause]
Image transferring from Ethernet has failed.
[Operation after Reply Key is touched]
For “Reply” cannot be done, the system must be shutdown.
-105
Fatal
OK #1:Fatal
Error
#2:CXDCAP.INI Open Error
#3:Install CXDCAP.INI file.
[Cause]
The cxdcap.ini file has not been installed.
[Operation after Reply Key is touched]
The system enters a fatal condition. All the sensor units will
therefore be disconnected. A service person needs to be called
in. However, you can transfer the image and use the study list.
-97
No cxdcap.ini file is found.
- 2 -
Summary of Contents for CXDI-50G
Page 20: ...2 Installation Manual 3 CXDI 50G Power Box assemble package 4 Grid optional 6 ...
Page 32: ...2 Installation Manual When normal imaging When timeout due to RX_REQ not negating 18 ...
Page 58: ...2 Installation Manual 44 Input parameters Fig 3 ...
Page 104: ...2 Installation Manual 90 2 Power Box Unit mm Mass 4 2Kg ...