1. General
5
Item Content
Remarks
Imaging restriction
(Imaging Prohibition)
When the internal temperature of
Imaging Unit is above 49 degree
Celsius, its state is changed to sleep
mode. And the Imaging prohibition
will be continued when the internal
temperature is below 48 degree
Celsius.
INDICATIO
N
BUSY SENSOR POWER
Color Orange
Green
Blue
Imaging unit
is off
Off Off Off
Imaging unit
is on
-- Off On
Preparing
imaging
-- Blinking
*1
On
Imaging
preparation
complete
-- Off On
Error status
-- Blinking
*2
On
Communi-
cating
Blinking
*3
--
On
Initialization
(when
startup)
-- Blinking
*4
On
Network not
set (when
startup)
-- Blinking
*5
On
User interface
Single type LED
Off: Imaging unit power is off
On in orange: Imaging unit power is
on (unable to perform imaging)
Blinking in green: Preparing
imaging/error status
On in green: Imaging preparation is
complete
*1: Turns on and off for 0.5 seconds each
*2: Turns on and off twice for 0.5 seconds, then
turns off for 0.5 seconds
*3: Turns on and off randomly
*4: Fades in for 1 second and fades out for 1
second
*5: Fades in for 2 seconds, and then turns off