9. Parameter Entries
87
Measurement parameters page 2
MEASUREMENT SETUP
2 / 2
Fraction Collector Settings...
Waste Valve
x
Waste Valve Settings...
Figure 30: Measurement parameters page 2
Fraction collector
settings...
Is displayed only if the item
Fraction collector
is enabled Upon selection
of this item you go to the page
Fraction collector settings
Waste valve
Tick this checkbox if a waste valve is employed.
Waste valve
settings...
Is displayed only if the item
Waste valve
is enabled. Upon selection of
this item you go to the page
Waste valve parameters
Fraction collector settings
FRACTION COLLECTOR SETTINGS
1 / 1
Controlled by:
Level
Vial Volumen [ml]:
10.0
Fill Time [s]:
2.0
Delay Time [s]:
1.0
Active from:
00:00:00
Active To:
00:00:00
On-Threshold:
100.0
Off-Threshold:
0.0
Time Window
Level Thresholds
Figure 31: Fraction collector settings
Controlled by
Control of the fraction collector either via
Threshold
or
Interval
.
Vial volume [ml]
Defines the vial volume.
Fill time [s]
Fill time for fractioning into vials (with broad peaks that have to be filled
into several vials). It is calculated by the system based on the vial volume
and the flow speed.
Delay time [s]
Time delay from the cell output to the fraction collector.