FlowStar2 LB 514
6. Validation using 3H- und 14C-Test sources
51
The instrument is in prerun mode. By pressing
<Start>
the validation
measurement starts and the current (upper right corner) as well as the in-
tegrated value (lower left corner) and the deviation from the nominal value
are displayed. If
„Prerun“ is still displayed after starting the measurement,
a prerun time is defined in the parameters. This prerun time is used to calm
down the test cell which could give higher values due to the exposure to
daylight.
After the prerun time the measurement starts automatically.
Once the measurement is finished either „PASSED“ or FAILED“ is dis-
played, depending if the results are within or outside the specified tolerance
level.
In ca
se of „FAILED“ the measurement can be repaeted by pressing
<Start>.
If the validation is „Passed“ the results can be saved pressing the
<Save>
button. A failed validation cannot be saved.
Menu
Menu
Pg
Start
5.1.2010
TIME
HH:mm:ss
ELAPSED
00:00:00
Current value cpm
Stat.error: 0.1%
62546
Deviation
nominal: 62856
0.5
4256.4
PRERUN
Validation
Esc
Start
Parameters
History
Start
5.1.2010
time
HH:mm:ss
elapsed
00:03:30
remain. 00:01:30
current. val. cpm
Stat. error 0.1
68611
deviation
nominal: 68958
0.1
4256.4
Measurement
Validation
Stop
Pg
Start
5.1.2010
cell
Test 14C
time
HH:mm:ss
elapsed 00:05:00
current val. cpm
stat. error: 0.0
69061
deviation
nominal: 68958
0.1
4256.4
Passed
Validation
Esc
Start
Save
Parameters
History