8. Structure and Operation of the FlowStar² Software
FlowStar2 LB 514
70
Info buttons
(see also chapter 15. Troubleshooting Check List)
<Status> button
Shows the general system status. Push the <
Status
> button to go to the
status page showing more information on the current system status, e.g.
which events have occurred.
We distinguish 5 states:
Ready
,
Error
,
Warning
,
Info
and
Measurement
.
Error
,
Warning
and
Info
: the respective text (below the line to the right of
the button) is displayed flashing inverse.
Warning
and
Error
: there is an additional acoustic signal.
Ready
display
The device is ready for measurement, no error/warning/info messages are
displayed.
Error
display
An error has been detected, the measurement was stopped completely.
Error events:
Leakage
Counting chamber open
Detector A+B HV off
Cell chip error
Detector error
Warning
display
A
warning
event
has
occurred,
no
errors
have
occurred.
Warnings
:
High luminescence
Overflow det. A
Overflow det. B
Info
display
An info event has occurred, no errors / warnings have occurred.
Infos:
System test overdue etc.
Measurement
display
A chromatogram measurement is active, no error/warning/info messages
are displayed.