10. Measurement
The device includes two measurement modes:
Continuous measurement: A continuous measurement starts automat-
ically after the device has been powered up using the measurement
parameters defined last (chapter 9.4). If you define parameters, for ex-
ample, the measurement continues to run in the background. The
measurement in the background is stopped temporarily only when the
energy
spectrum
is
called
in
the
service
functions.
This mode is useful when no external hardware is controlled through
the FlowStar² since this requires an external start signal.
Measurements over a limited time: After selection of a method and def-
inition of the measurement parameters, these measurements are
started on the
Measurement
menu and can be stopped again.
This measurement mode must be used when external hardware (scin-
tillator pump, waste valve etc.) is controlled though the FlowStar². It re-
quires either a manual start (touch screen) or an external start signal.
10.1 Measurement Menu
The currently measured values of both counting channels are displayed
graphically and numerically on the
Measurement
menu. Push
Pg
to
switch between display of the radioactivity and the analog signals.
Info row and info buttons
First row
The method and the run name are displayed alternatingly.
<Status>
Shows the current status. Push this button to view possible
error messages (see chapter 8.2 and 15).
<Cell>
Shows the currently used type of measuring cell. Push this
button to view the measuring cell parameters.
<LS pump>
This button is displayed only when the scintillator pump has
been enabled on the
Configuration
menu. The pump status
is indicated (either
ON
or
OFF
). Push this button to view the
pump parameters.
Status
READY
CelL
Test 14C
LS-Pump
OFF
TIME
4256.4
432.4
2 0
1 0
0
0
2
4
6
8
2 0
1 0
0
0
2
4
6
8
Menu
Pg
Start
Figure 33: Measurement menu