PN B05577AC
6-45
Sample Modules
Micro Liquid Module
6
Figure 6.31
Locator Notch
Filling Micro Liquid Module with Diluent
1
Open the door to the optical bench. If the sample cell is placed on the stand, remove it.
2
Remove the cap.
3
Tilt the cell toward the back (away from the locator pin) in order to trap less air inside the cup.
Pour fluid slowly into sample entry (
).
Figure 6.32
Cell Diluent / Sample Entry
1.
Cell
2.
Locator Pin
3.
Locator Notch