PN B05577AC
6-2
Sample Modules
Aqueous Liquid Module
ALM Description
System Description
The ALM system consists of:
•
A sample vessel containing the suspension fluid and dispersed sample particles.
•
A sample cell.
•
A sonicator that aids in the dispersion of the sample.
•
A variable speed circulation pump to circulate the particles through the sample cell.
•
Inlet and drain hoses.
•
Hoses to transfer fluid from the vessel to the sample cell.
•
Liquid level sensors.
The ALM is intended to be used with water only and holds a total volume of 1250 mL. Using the
automatic fill mode, the system only uses 1100 mL. The ALM has the capability of auto-filling and
auto-rinsing using a self-contained internal pump that is operated through software commands.
The pump fills the vessel to a set level controlled by liquid level sensors (refer to
). If the
vessel is manually filled, an alarm will sound to indicate when the liquid level has reached the
overflow sensor, this is intended to avoid fluid overflow.
Figure 6.2
Liquid Level Sensors
Indicators
A set of LEDs placed in the front panel of the module is used to provide visual information about the
status of the system when it is operated in manual mode.
shows these LEDs and their
functions. The indicators are active only in manual mode though the functions indicated can also
be accessed through the software.
1.
Liquid Level Sensors
Summary of Contents for LS 13 320
Page 4: ...PN B05577AC iv Revision History ...
Page 12: ...PN B05577AC xii Safety Notice Scope of Manual ...
Page 30: ...PN B05577AC xxx LS 13 320 Introduction System Components ...
Page 57: ...PN B05577AC 2 21 Installation Making Measurements 2 Figure 2 18 Run Cycle Options Dialog ...
Page 59: ...PN B05577AC 3 1 CHAPTER 3 LS 13 320 Software Figure 3 1 LS 13 320 Software ...
Page 80: ...PN B05577AC 3 22 LS 13 320 Software Preference Options ...
Page 86: ...PN B05577AC 3 28 LS 13 320 Software Preference Options ...
Page 116: ...PN B05577AC 4 10 Regulatory Compliance 21 CFR Part 11 Starting Security Enabled Software ...
Page 169: ...PN B05577AC 6 39 Sample Modules Micro Liquid Module 6 Figure 6 27 SOM List Dialog ...
Page 216: ...PN B05577AC 6 86 Sample Modules Universal Liquid Module ...
Page 228: ...PN B05577AC B 10 Sample Handling Diluent Selection ...
Page 238: ...PN B05577AC C 10 Optical Models Statistics ...
Page 252: ...PN B05577AC Warranty 2 Beckman Coulter Inc Customer End User License Agreement ...
Page 253: ......