PN B05577AC
6-85
Sample Modules
Universal Liquid Module
6
ULM Troubleshooting
ULM Specifications
Problem
Possible Cause/Solution
No communication with optical bench
•
Module docked incorrectly. Eject module and re-dock. Apply
slight pressure on the front of module while docking.
Module will not fill
•
Check to see if it is already full
•
Check that connections are air-tight
•
Check that the fill hose is submerged in the diluent.
•
The valve should make an audible sound. If it does not,
contact Beckman Coulter service.
Module will not drain
•
Check that the waste container is not full.
•
Check that the drain hose is clear of obstructions.
Pump will not work
•
Module is not completely full of fluid.
•
Check power to the system.
•
Check belt drive.
Module leaking
•
Check fittings.
•
If leakage is inside, contact Beckman Coulter
•
service.
Item
Specification
Power
The ULM derives its power directly from the optical bench < 50 W peak power
Temperature
10° to 40° C (50° to 104° F)
Humidity
0 to 90% without condensation
Dimensions
Height
49.5 cm (19.5 in)
Width
21.0 cm (8.25 in)
Depth
a
11.4 cm (4.5 in)
a. Visible portion of module once docked into the optical bench
Depth
b
b. Includes optical cell
Weight
8.85 kg (19.5 lb)
Summary of Contents for LS 13 320
Page 4: ...PN B05577AC iv Revision History ...
Page 12: ...PN B05577AC xii Safety Notice Scope of Manual ...
Page 30: ...PN B05577AC xxx LS 13 320 Introduction System Components ...
Page 57: ...PN B05577AC 2 21 Installation Making Measurements 2 Figure 2 18 Run Cycle Options Dialog ...
Page 59: ...PN B05577AC 3 1 CHAPTER 3 LS 13 320 Software Figure 3 1 LS 13 320 Software ...
Page 80: ...PN B05577AC 3 22 LS 13 320 Software Preference Options ...
Page 86: ...PN B05577AC 3 28 LS 13 320 Software Preference Options ...
Page 116: ...PN B05577AC 4 10 Regulatory Compliance 21 CFR Part 11 Starting Security Enabled Software ...
Page 169: ...PN B05577AC 6 39 Sample Modules Micro Liquid Module 6 Figure 6 27 SOM List Dialog ...
Page 216: ...PN B05577AC 6 86 Sample Modules Universal Liquid Module ...
Page 228: ...PN B05577AC B 10 Sample Handling Diluent Selection ...
Page 238: ...PN B05577AC C 10 Optical Models Statistics ...
Page 252: ...PN B05577AC Warranty 2 Beckman Coulter Inc Customer End User License Agreement ...
Page 253: ......