PN B05577AC
6-6
Sample Modules
Aqueous Liquid Module
2
In the auto-dock tray, you will notice stainless steel pins extending into the center tray area
both at the back of the tray and in the front of the tray. These pins fit into slots on the ALM
Sample Cell. The back pins prevent the module from tipping out and the forward pins locate the
module in the tray.
3
Hold the ALM Sample Cell by the handle and place it in the auto-dock tray by tipping the Sample
Cell to the rear at an approximately 30 degree angle into the auto-dock tray. This is done to clear
the forward pins and engage the back pins first.
4
When these pins have engaged into the slots on the back of the Sample Cell, tip the Sample Cell
up to its vertical position. The Sample Cell should now be resting on the auto-dock tray.
5
Press the
EJECT MODULE
button and the Sample Cell will be drawn into the LS 13 320.
CAUTION
Do not place fingers inside bench as the sample cell docks into bench.
CAUTION
Do not place fingers inside bench as automatic sliding door closes.
Software
This section covers the controls found in the software program to operate the ALM. It also covers
the parts of the Wizards (Standard Operating Methods (SOMs)) that deal with the ALM.
Summary of Contents for LS 13 320
Page 4: ...PN B05577AC iv Revision History ...
Page 12: ...PN B05577AC xii Safety Notice Scope of Manual ...
Page 30: ...PN B05577AC xxx LS 13 320 Introduction System Components ...
Page 57: ...PN B05577AC 2 21 Installation Making Measurements 2 Figure 2 18 Run Cycle Options Dialog ...
Page 59: ...PN B05577AC 3 1 CHAPTER 3 LS 13 320 Software Figure 3 1 LS 13 320 Software ...
Page 80: ...PN B05577AC 3 22 LS 13 320 Software Preference Options ...
Page 86: ...PN B05577AC 3 28 LS 13 320 Software Preference Options ...
Page 116: ...PN B05577AC 4 10 Regulatory Compliance 21 CFR Part 11 Starting Security Enabled Software ...
Page 169: ...PN B05577AC 6 39 Sample Modules Micro Liquid Module 6 Figure 6 27 SOM List Dialog ...
Page 216: ...PN B05577AC 6 86 Sample Modules Universal Liquid Module ...
Page 228: ...PN B05577AC B 10 Sample Handling Diluent Selection ...
Page 238: ...PN B05577AC C 10 Optical Models Statistics ...
Page 252: ...PN B05577AC Warranty 2 Beckman Coulter Inc Customer End User License Agreement ...
Page 253: ......