Tornado Dry Powder System
Loading an SOP
To load an SOP:
1
Run
.
2
Select Load an SOP. SOPs can also be loaded from the status panel by clicking on the
Load SOP
button.
3
Select the desired SOP file from the SOP subfolder or the subfolder where the file was last saved.
Cleaning the Tornado DPS
It is important for all glass surfaces in your sample system to be clean, as dusty or coated optical
surfaces can cause erroneous results. Some samples may coat the inside of the sample cell windows.
If the instrument is in a dusty or smoky environment or if vapors are present, the lenses and outer
surfaces of the windows may become coated. Check the optical surfaces for cleanliness at least every
three months, and clean if needed. Depending on the environment, cleaning may be required more
frequently.
Using the Cleaning Sample
The sample cell and tubing may be cleaned by using the provided cleaning sample. Load
approximately 15 cc to 25 cc of the cleaning sample into the sample holder and select the
Run
Cleaner
option from the Control menu.
Summary of Contents for LS 13 320
Page 4: ...PN B05577AC iv Revision History ...
Page 12: ...PN B05577AC xii Safety Notice Scope of Manual ...
Page 30: ...PN B05577AC xxx LS 13 320 Introduction System Components ...
Page 57: ...PN B05577AC 2 21 Installation Making Measurements 2 Figure 2 18 Run Cycle Options Dialog ...
Page 59: ...PN B05577AC 3 1 CHAPTER 3 LS 13 320 Software Figure 3 1 LS 13 320 Software ...
Page 80: ...PN B05577AC 3 22 LS 13 320 Software Preference Options ...
Page 86: ...PN B05577AC 3 28 LS 13 320 Software Preference Options ...
Page 116: ...PN B05577AC 4 10 Regulatory Compliance 21 CFR Part 11 Starting Security Enabled Software ...
Page 169: ...PN B05577AC 6 39 Sample Modules Micro Liquid Module 6 Figure 6 27 SOM List Dialog ...
Page 216: ...PN B05577AC 6 86 Sample Modules Universal Liquid Module ...
Page 228: ...PN B05577AC B 10 Sample Handling Diluent Selection ...
Page 238: ...PN B05577AC C 10 Optical Models Statistics ...
Page 252: ...PN B05577AC Warranty 2 Beckman Coulter Inc Customer End User License Agreement ...
Page 253: ......