PN B05577AC
6-47
Sample Modules
Micro Liquid Module
6
Cleaning the MLM
IMPORTANT
Always remove the Micro Liquid Cell from the sample stand to add fluid or sample in order to
prevent spills on the Micro Liquid Sample Stand or inside the optical module. If you follow this advice, only
the Micro Liquid Cell will require occasional cleaning.
Disassemble the cell
.
1
Unscrew the cell cap.
2
Pour the fluid out of the cell.
3
Remove the O-ring from the top of the cell.
4
Unscrew the cell body from the cup and remove the stirrer bar.
5
Remove the O-rings from the stirring bar.
Clean the cell components
1
Wash everything except the cell body in a 2% solution of MICRO Critical lab cleaner. Rinse and
dry.
2
Clean the inside of the cell body by rinsing with acetone or isopropyl alcohol.
IMPORTANT
Do not use any abrasive cleaners in the cell.
3
If any particles or dirt remain on the inside of the cell windows after rinsing, use lens cleaning
fluid and the swab to clean the windows. Clean the windows with a circular motion.
4
Clean the outside of the cell windows with lens cleaning fluid and lens tissue using a circular
motion.
Summary of Contents for LS 13 320
Page 4: ...PN B05577AC iv Revision History ...
Page 12: ...PN B05577AC xii Safety Notice Scope of Manual ...
Page 30: ...PN B05577AC xxx LS 13 320 Introduction System Components ...
Page 57: ...PN B05577AC 2 21 Installation Making Measurements 2 Figure 2 18 Run Cycle Options Dialog ...
Page 59: ...PN B05577AC 3 1 CHAPTER 3 LS 13 320 Software Figure 3 1 LS 13 320 Software ...
Page 80: ...PN B05577AC 3 22 LS 13 320 Software Preference Options ...
Page 86: ...PN B05577AC 3 28 LS 13 320 Software Preference Options ...
Page 116: ...PN B05577AC 4 10 Regulatory Compliance 21 CFR Part 11 Starting Security Enabled Software ...
Page 169: ...PN B05577AC 6 39 Sample Modules Micro Liquid Module 6 Figure 6 27 SOM List Dialog ...
Page 216: ...PN B05577AC 6 86 Sample Modules Universal Liquid Module ...
Page 228: ...PN B05577AC B 10 Sample Handling Diluent Selection ...
Page 238: ...PN B05577AC C 10 Optical Models Statistics ...
Page 252: ...PN B05577AC Warranty 2 Beckman Coulter Inc Customer End User License Agreement ...
Page 253: ......