PN B05577AC
E-3
Maintenance
Fourier Lenses
E
7
Repeat steps
through
to clean the lower Fourier lens.
CAUTION
If the lens or soft anti-reflective coating is scratched, the lens must be replaced.
Call your local Beckman Coulter Representative.
8
Carefully inspect the lens' surface for cleanliness by shining a bright light on it and looking at
the surface from all angles.
9
Dock the sample module back into the instrument.
10
Select the
Measure Offsets
,
Auto-align
and
Measure Background
functions in the Run Cycle
dialog box before running your next sample.
Projection Lens
Use this procedure to clean the projection lens. To clean the projection lens:
1
If a sample module is docked, undock it, if not, open the Optical Module's door.
CAUTION
Use lens tissues only once, then discard. Never use siliconecoated, eyeglass lens
tissues to clean lenses. They can leave a film. Wash your hands thoroughly before
cleaning any lens or sample cell window. Do not touch optical surfaces with your
fingers or skin. Body oils are difficult to remove from lenses without harming the
anti-reflective coating.
2
Fold two or three lens tissues in half.
3
Squirt lens cleaner onto the folded lens tissues.
4
Place your finger in back of the lens tissue(s).
5
Starting at the outermost edge of the component, wipe in a spiral pattern to the middle. Refer
to
Summary of Contents for LS 13 320
Page 4: ...PN B05577AC iv Revision History ...
Page 12: ...PN B05577AC xii Safety Notice Scope of Manual ...
Page 30: ...PN B05577AC xxx LS 13 320 Introduction System Components ...
Page 57: ...PN B05577AC 2 21 Installation Making Measurements 2 Figure 2 18 Run Cycle Options Dialog ...
Page 59: ...PN B05577AC 3 1 CHAPTER 3 LS 13 320 Software Figure 3 1 LS 13 320 Software ...
Page 80: ...PN B05577AC 3 22 LS 13 320 Software Preference Options ...
Page 86: ...PN B05577AC 3 28 LS 13 320 Software Preference Options ...
Page 116: ...PN B05577AC 4 10 Regulatory Compliance 21 CFR Part 11 Starting Security Enabled Software ...
Page 169: ...PN B05577AC 6 39 Sample Modules Micro Liquid Module 6 Figure 6 27 SOM List Dialog ...
Page 216: ...PN B05577AC 6 86 Sample Modules Universal Liquid Module ...
Page 228: ...PN B05577AC B 10 Sample Handling Diluent Selection ...
Page 238: ...PN B05577AC C 10 Optical Models Statistics ...
Page 252: ...PN B05577AC Warranty 2 Beckman Coulter Inc Customer End User License Agreement ...
Page 253: ......