Sample Handling
Material Handling
A representative sample can be obtained by implementing one of the five common sampling
procedures. The five procedures are listed in descending order with respect to their efficiency at
getting good sample representation.
Spinning Riffler
The spinning riffler consists of a rotary sample tray, hopper, and a vibratory feeder. The sample fills
the mass flow hopper without heaping in piles and falls from the hopper to the chute of the
vibratory feeder. The feeder supplies a constant flow rate to the collection boxes, which move in a
circular motion.This method, according to statistics, yields the most accurate results. It is
recommended that samples be subdivided to the amount of a gram and the sample must be free
flowing.
Figure B.1
Figure 1. Spinning Riffler
Chute splitter
The sample is placed in a V-shaped hopper and migrates down a series of chutes that alternately
feed two trays placed on either side of the trough. The sample is poured into the hopper and
repeatedly halved until a desired sample amount is obtained. Satisfactory sample division can be
obtained only if the operator loads the sample into the trough carefully without segregation
occurring.
Summary of Contents for LS 13 320
Page 4: ...PN B05577AC iv Revision History ...
Page 12: ...PN B05577AC xii Safety Notice Scope of Manual ...
Page 30: ...PN B05577AC xxx LS 13 320 Introduction System Components ...
Page 57: ...PN B05577AC 2 21 Installation Making Measurements 2 Figure 2 18 Run Cycle Options Dialog ...
Page 59: ...PN B05577AC 3 1 CHAPTER 3 LS 13 320 Software Figure 3 1 LS 13 320 Software ...
Page 80: ...PN B05577AC 3 22 LS 13 320 Software Preference Options ...
Page 86: ...PN B05577AC 3 28 LS 13 320 Software Preference Options ...
Page 116: ...PN B05577AC 4 10 Regulatory Compliance 21 CFR Part 11 Starting Security Enabled Software ...
Page 169: ...PN B05577AC 6 39 Sample Modules Micro Liquid Module 6 Figure 6 27 SOM List Dialog ...
Page 216: ...PN B05577AC 6 86 Sample Modules Universal Liquid Module ...
Page 228: ...PN B05577AC B 10 Sample Handling Diluent Selection ...
Page 238: ...PN B05577AC C 10 Optical Models Statistics ...
Page 252: ...PN B05577AC Warranty 2 Beckman Coulter Inc Customer End User License Agreement ...
Page 253: ......