PN B05577AC
E-4
Maintenance
Fourier Lenses
Figure E.2
Projection Lens Location and Cleaning Technique
6
Discard the used lens tissue(s).
CAUTION
If the lens or soft anti-reflective coating is scratched, the lens must be replaced.
Call your local Beckman Coulter Representative.
7
Carefully inspect the lens surface for cleanliness by shining a bright light on it and looking at
the surface from all angles.
8
Dock the sample module back into the instrument.
9
Select the
Measure Offsets
,
Auto-align
and
Measure Background
functions in the Run Cycle
dialog box before running your next sample.
1.
Projection Lens
2.
Tissue
Summary of Contents for LS 13 320
Page 4: ...PN B05577AC iv Revision History ...
Page 12: ...PN B05577AC xii Safety Notice Scope of Manual ...
Page 30: ...PN B05577AC xxx LS 13 320 Introduction System Components ...
Page 57: ...PN B05577AC 2 21 Installation Making Measurements 2 Figure 2 18 Run Cycle Options Dialog ...
Page 59: ...PN B05577AC 3 1 CHAPTER 3 LS 13 320 Software Figure 3 1 LS 13 320 Software ...
Page 80: ...PN B05577AC 3 22 LS 13 320 Software Preference Options ...
Page 86: ...PN B05577AC 3 28 LS 13 320 Software Preference Options ...
Page 116: ...PN B05577AC 4 10 Regulatory Compliance 21 CFR Part 11 Starting Security Enabled Software ...
Page 169: ...PN B05577AC 6 39 Sample Modules Micro Liquid Module 6 Figure 6 27 SOM List Dialog ...
Page 216: ...PN B05577AC 6 86 Sample Modules Universal Liquid Module ...
Page 228: ...PN B05577AC B 10 Sample Handling Diluent Selection ...
Page 238: ...PN B05577AC C 10 Optical Models Statistics ...
Page 252: ...PN B05577AC Warranty 2 Beckman Coulter Inc Customer End User License Agreement ...
Page 253: ......