PN B05577AC
6-18
Sample Modules
Aqueous Liquid Module
Figure 6.10
Diffraction Window
5
Cup your hand over the loose retaining ring and tilt the sample cell stand until the retaining
ring falls into your hand.
NOTE
You can now lay down the sample stand so the window to be replaced is facing up.
6
Stick a piece of masking tape on the window and lift the tape up to remove the O-ring and
window. This may take several tries.
7
Discard the old O-ring, window and nylon screws.
8
Hold the new window by its edge to determine the coated side. Refer to the figure below to
determine which side is the correct one.
1.
Difraction Cell Window
2.
O-ring
3.
Retaining Ring
4.
Nylon Screws
5.
Sample Cell
6.
Supply Hose
Summary of Contents for LS 13 320
Page 4: ...PN B05577AC iv Revision History ...
Page 12: ...PN B05577AC xii Safety Notice Scope of Manual ...
Page 30: ...PN B05577AC xxx LS 13 320 Introduction System Components ...
Page 57: ...PN B05577AC 2 21 Installation Making Measurements 2 Figure 2 18 Run Cycle Options Dialog ...
Page 59: ...PN B05577AC 3 1 CHAPTER 3 LS 13 320 Software Figure 3 1 LS 13 320 Software ...
Page 80: ...PN B05577AC 3 22 LS 13 320 Software Preference Options ...
Page 86: ...PN B05577AC 3 28 LS 13 320 Software Preference Options ...
Page 116: ...PN B05577AC 4 10 Regulatory Compliance 21 CFR Part 11 Starting Security Enabled Software ...
Page 169: ...PN B05577AC 6 39 Sample Modules Micro Liquid Module 6 Figure 6 27 SOM List Dialog ...
Page 216: ...PN B05577AC 6 86 Sample Modules Universal Liquid Module ...
Page 228: ...PN B05577AC B 10 Sample Handling Diluent Selection ...
Page 238: ...PN B05577AC C 10 Optical Models Statistics ...
Page 252: ...PN B05577AC Warranty 2 Beckman Coulter Inc Customer End User License Agreement ...
Page 253: ......