PN B05577AC
6-32
Sample Modules
Micro Liquid Module
Loading the MLM
To install the MLM:
1
Open the LS 13 320 door to load the sample cell by undocking the auto-docking tray. Press the
OPEN
button on the instrument and the sliding door moves to the left. When it is open, press
the
EJECT MODULE
button to extend the tray toward the user.
2
In the auto-dock tray, you will notice stainless steel pins extending into the center tray area
both at the back of the tray and in the front of the tray. These pins fit into slots on the MLM. The
back pins prevent the module from tipping out and the forward pins secure the module in the
tray.
3
Hold the MLM by the top with one hand and on the bottom with the other hand. Place it in the
auto-dock tray by tipping the Sample Cell to the rear at an approximately 30 degree angle into
the auto-dock tray. This is done to clear the forward pins and engage the back pins first.
4
When these pins have engaged into the slots on the back of the MLM, tip the Sample Cell up to
its vertical position. The MLM should now be resting on the auto-dock tray.
5
Press the
EJECT MODULE
button and the MLM will be drawn into the LS 13 320.
CAUTION
Do not place fingers inside bench as module docks into bench.
CAUTION
Do not place fingers inside bench as automatic sliding door closes.
Software
This section covers the Wizards (Standard Operating Methods (SOMs)) and controls found in the
software program to operate the Micro Liquid Module.
Control Menu
The MLM system can be operated through a series of manual controls accessed through the
software.
Summary of Contents for LS 13 320
Page 4: ...PN B05577AC iv Revision History ...
Page 12: ...PN B05577AC xii Safety Notice Scope of Manual ...
Page 30: ...PN B05577AC xxx LS 13 320 Introduction System Components ...
Page 57: ...PN B05577AC 2 21 Installation Making Measurements 2 Figure 2 18 Run Cycle Options Dialog ...
Page 59: ...PN B05577AC 3 1 CHAPTER 3 LS 13 320 Software Figure 3 1 LS 13 320 Software ...
Page 80: ...PN B05577AC 3 22 LS 13 320 Software Preference Options ...
Page 86: ...PN B05577AC 3 28 LS 13 320 Software Preference Options ...
Page 116: ...PN B05577AC 4 10 Regulatory Compliance 21 CFR Part 11 Starting Security Enabled Software ...
Page 169: ...PN B05577AC 6 39 Sample Modules Micro Liquid Module 6 Figure 6 27 SOM List Dialog ...
Page 216: ...PN B05577AC 6 86 Sample Modules Universal Liquid Module ...
Page 228: ...PN B05577AC B 10 Sample Handling Diluent Selection ...
Page 238: ...PN B05577AC C 10 Optical Models Statistics ...
Page 252: ...PN B05577AC Warranty 2 Beckman Coulter Inc Customer End User License Agreement ...
Page 253: ......