PN B05577AC
6-11
Sample Modules
Aqueous Liquid Module
6
6
In Step 4 the run length is chosen. For analyses that require collection of PIDS data, a run length
of at least 90 seconds is recommended. For analyses without PIDS, 60 seconds is the
recommended time for the analysis, though analyses may be as short as 10 seconds. The optical
model can be selected during this step. For more on optical models see
. The
Pump Speed
can be adjusted in this dialog. It is recommended that samples
requiring some type of surfactant, that pump speeds be kept low, i.e. 30%, to prevent the
formation of air bubbles. Should bubbles be present, alternating high pump speed with low
pump speed is useful when trying to de-bubble the suspending fluid. If the pump speed is
increased to its maximum (100%) and then reduced to an approximate 50%, air bubbles can be
eliminated. This procedure may need to be repeated two to three times to eliminate the
bubbles. High pump speeds are also useful when measuring large, dense particles. You can
access the manual speed control from the Control menu by selecting
Pump Speed
.
The
Directory
option allows the user to select a custom directory to save the data under or to
simply utilize the current directory for saving data.
Summary of Contents for LS 13 320
Page 4: ...PN B05577AC iv Revision History ...
Page 12: ...PN B05577AC xii Safety Notice Scope of Manual ...
Page 30: ...PN B05577AC xxx LS 13 320 Introduction System Components ...
Page 57: ...PN B05577AC 2 21 Installation Making Measurements 2 Figure 2 18 Run Cycle Options Dialog ...
Page 59: ...PN B05577AC 3 1 CHAPTER 3 LS 13 320 Software Figure 3 1 LS 13 320 Software ...
Page 80: ...PN B05577AC 3 22 LS 13 320 Software Preference Options ...
Page 86: ...PN B05577AC 3 28 LS 13 320 Software Preference Options ...
Page 116: ...PN B05577AC 4 10 Regulatory Compliance 21 CFR Part 11 Starting Security Enabled Software ...
Page 169: ...PN B05577AC 6 39 Sample Modules Micro Liquid Module 6 Figure 6 27 SOM List Dialog ...
Page 216: ...PN B05577AC 6 86 Sample Modules Universal Liquid Module ...
Page 228: ...PN B05577AC B 10 Sample Handling Diluent Selection ...
Page 238: ...PN B05577AC C 10 Optical Models Statistics ...
Page 252: ...PN B05577AC Warranty 2 Beckman Coulter Inc Customer End User License Agreement ...
Page 253: ......