Sample Modules
Auto-Prep Station
6
ALM Troubleshooting
ALM Specifications
Auto-Prep Station
The Auto-Prep Station (APS) is intended to be used with the LS 13 320 optical bench and the
Aqueous Liquid Module (ALM). It is capable of adding dispersant as well as sonicating a total of thirty
samples. It is intended for use with aqueous samples only.
Problem
Possible Cause/Solution
No communication with optical bench
•
Module docked incorrectly.
•
Undock module and re-dock.
Module won't fill
•
Check to see if it is already full.
•
Check that connections are air-tight.
•
Make sure there is enough inlet pressure.
Module won't drain
•
Check that drain hose is clear of obstructions.
Pump will not work
•
Module is not completely filled with fluid.
•
Check power to the system.
•
Sample cell not fully docked.
Module leaking
•
Check fittings.
•
If leakage is inside, contact Beckman Coulter service.
Item
Specification
Power
The ALM derives its power directly from the optical bench < 50 W peak-power
Temperature
10° to 40°C (50° to 104°F)
Humidity
0 to 90% without condensation
Dimensions
Height
29.85 cm (11.75 in)
Width
38.7 cm (15.25 in)
Depth
34.9 cm (13.75 in)
Weight
15.4 kg (34.0 lb)
Summary of Contents for LS 13 320
Page 4: ...PN B05577AC iv Revision History ...
Page 12: ...PN B05577AC xii Safety Notice Scope of Manual ...
Page 30: ...PN B05577AC xxx LS 13 320 Introduction System Components ...
Page 57: ...PN B05577AC 2 21 Installation Making Measurements 2 Figure 2 18 Run Cycle Options Dialog ...
Page 59: ...PN B05577AC 3 1 CHAPTER 3 LS 13 320 Software Figure 3 1 LS 13 320 Software ...
Page 80: ...PN B05577AC 3 22 LS 13 320 Software Preference Options ...
Page 86: ...PN B05577AC 3 28 LS 13 320 Software Preference Options ...
Page 116: ...PN B05577AC 4 10 Regulatory Compliance 21 CFR Part 11 Starting Security Enabled Software ...
Page 169: ...PN B05577AC 6 39 Sample Modules Micro Liquid Module 6 Figure 6 27 SOM List Dialog ...
Page 216: ...PN B05577AC 6 86 Sample Modules Universal Liquid Module ...
Page 228: ...PN B05577AC B 10 Sample Handling Diluent Selection ...
Page 238: ...PN B05577AC C 10 Optical Models Statistics ...
Page 252: ...PN B05577AC Warranty 2 Beckman Coulter Inc Customer End User License Agreement ...
Page 253: ......