PN B05577AC
6-78
Sample Modules
Universal Liquid Module
Figure 6.43
Miscibility Guide
Cleaning the ULM
It is important for all glass surfaces in your sample system to be clean, as dusty or coated optical
surfaces can cause erroneous results. Some samples or suspension fluids may coat the inside of the
sample cell windows. If the instrument is in a dusty or smoky environment or if vapors are present,
the lenses and outer surfaces of the windows may become coated. Check the optical surfaces for
cleanliness at least every three months, and clean if needed. Depending on the environment,
cleaning may be required more frequently.
Cleaning Fluids
Beckman Coulter recommends using the cleaning solution supplied with the module for cleaning
the inner surfaces of the ULM sample system. Beckman Coulter recommends lens cleaning paper
for cleaning the exterior surfaces of the lenses and sample cell windows.
Summary of Contents for LS 13 320
Page 4: ...PN B05577AC iv Revision History ...
Page 12: ...PN B05577AC xii Safety Notice Scope of Manual ...
Page 30: ...PN B05577AC xxx LS 13 320 Introduction System Components ...
Page 57: ...PN B05577AC 2 21 Installation Making Measurements 2 Figure 2 18 Run Cycle Options Dialog ...
Page 59: ...PN B05577AC 3 1 CHAPTER 3 LS 13 320 Software Figure 3 1 LS 13 320 Software ...
Page 80: ...PN B05577AC 3 22 LS 13 320 Software Preference Options ...
Page 86: ...PN B05577AC 3 28 LS 13 320 Software Preference Options ...
Page 116: ...PN B05577AC 4 10 Regulatory Compliance 21 CFR Part 11 Starting Security Enabled Software ...
Page 169: ...PN B05577AC 6 39 Sample Modules Micro Liquid Module 6 Figure 6 27 SOM List Dialog ...
Page 216: ...PN B05577AC 6 86 Sample Modules Universal Liquid Module ...
Page 228: ...PN B05577AC B 10 Sample Handling Diluent Selection ...
Page 238: ...PN B05577AC C 10 Optical Models Statistics ...
Page 252: ...PN B05577AC Warranty 2 Beckman Coulter Inc Customer End User License Agreement ...
Page 253: ......